Abnormally high Patient State Index associated with epicardial pacing: a case report

  • Hiroki Yonezawa
  • , Takuma Maeda
  • , Yoshiaki Takise
  • , Masahiro Morinaga
  • , Yoshihiko Ohnishi

Research output: Contribution to journal › Article › peer-review

Abstract

Background: Anesthesiologists monitor electroencephalography (EEG) intraoperatively to maintain adequate depth of anesthesia. However, the EEG signal is affected by noise and interference. The SedLine® is a brain function monitor with which the Patient State Index (PSI) is calculated. In this study, we report abnormally high PSI values associated with epicardial pacing during open heart surgery. Case presentation: A 50-year-old man was scheduled for total arch replacement. Atrial demand pacing was started before weaning from cardiopulmonary bypass. The PSI increased from 30 to 80 soon after the start of pacing, and the EEG waveform showed spikes synchronized with the pacing. As the pacing output was lowered, the spikes on the EEG attenuated and disappeared, and the PSI decreased to < 40. When the pacing output was increased again, the spikes recurred, and the PSI increased again. Conclusions: Pacemaker spikes may cause contamination of the EEG, resulting in abnormally high PSI values.

Original languageEnglish
Article number35
JournalJA Clinical Reports
Volume8
Issue number1
DOIs
Publication statusPublished - 2022 Dec
Externally publishedYes

Keywords

  • Electroencephalography
  • Monitoring
  • Patient State Index

ASJC Scopus subject areas

  • Anesthesiology and Pain Medicine

Fingerprint

Dive into the research topics of 'Abnormally high Patient State Index associated with epicardial pacing: a case report'. Together they form a unique fingerprint.

Cite this