Abstract
In this study, an automatic measurement of HNO3 was investigated by using a diffusion scrubber and ion chromatography. The collection efficiency of HNO3 standard gas of 36 μg/m3 was 93.1°3.3% (n=4) by using 50 cm lengh of the diffusion scrubber at 1 2/min of air flow. The lower detection limit of HN03 by this method was 0.01 μg/m3 at sampling volume of 60 60 The concentration of HNO3 in the urban atmosphere is a few μg/m3 It is suitable for this method to measure HNO3 at ambient level.
Original language | English |
---|---|
Pages (from-to) | 1025-1028 |
Number of pages | 4 |
Journal | analytical sciences |
Volume | 7 |
DOIs | |
Publication status | Published - 1991 Jan |
Externally published | Yes |
Keywords
- diffusion scrubber
- gaseous nitric acid
- ion chromatography
- porous polytetra-fluoroethylene(PPTFE) tube
ASJC Scopus subject areas
- Analytical Chemistry