TY - JOUR
T1 - Application of real-time radiation dosimetry using a new silicon LET sensor
AU - Doke, Tadayoshi
AU - Hayashi, Takayoshi
AU - Kikuchi, Jun
AU - Nagaoka, Shunji
AU - Nakano, T.
AU - Sakaguchi, T.
AU - Terasawa, K.
AU - Badhwar, G. D.
N1 - Copyright:
Copyright 2007 Elsevier B.V., All rights reserved.
PY - 1999/12/6
Y1 - 1999/12/6
N2 - A new type of real-time radiation monitoring device, RRMD-III, consisting of three double-sided silicon strip detectors (DSSDs), has been developed and tested on-board the Space Shuttle mission STS-84. The test succeeded in measuring the linear energy transfer (LET) distribution over the range of 0.2 keV/μm to 600 keV/μm for 178 h. The Shuttle cruised at an altitude of 300 to 400 km and an inclination angle of 51.6°for 221.3 h, which is equivalent to the International Space Station orbit. The LET distribution obtained for particles was investigated by separating it into galactic cosmic ray (GCR) particles and trapped particles in the South Atlantic Anomaly (SAA) region. The result shows that the contribution in dose-equivalent due to GCR particles is almost equal to that from trapped particles. The total absorbed dose rate during the mission was 0.611 mGy/day; the effective quality factor, 1.64; and the dose equivalent rate, 0.998 mSv/day. The average absorbed dose rates are 0.158 mGy/min for GCR particles and 3.67 mGy/min for trapped particles. The effective quality factors are 2.48 for GCR particles and 1.19 for trapped particles. The absorbed doses obtained by the RRMD-III and a conventional method using TLD (Mg2SiO4), which was placed around the RRMD-III were compared. It was found that the TLDs showed a lower efficiency, just 58% of absorbed dose registered by the RRMD-III. Copyright (C) 1999 Elsevier Science B.V.
AB - A new type of real-time radiation monitoring device, RRMD-III, consisting of three double-sided silicon strip detectors (DSSDs), has been developed and tested on-board the Space Shuttle mission STS-84. The test succeeded in measuring the linear energy transfer (LET) distribution over the range of 0.2 keV/μm to 600 keV/μm for 178 h. The Shuttle cruised at an altitude of 300 to 400 km and an inclination angle of 51.6°for 221.3 h, which is equivalent to the International Space Station orbit. The LET distribution obtained for particles was investigated by separating it into galactic cosmic ray (GCR) particles and trapped particles in the South Atlantic Anomaly (SAA) region. The result shows that the contribution in dose-equivalent due to GCR particles is almost equal to that from trapped particles. The total absorbed dose rate during the mission was 0.611 mGy/day; the effective quality factor, 1.64; and the dose equivalent rate, 0.998 mSv/day. The average absorbed dose rates are 0.158 mGy/min for GCR particles and 3.67 mGy/min for trapped particles. The effective quality factors are 2.48 for GCR particles and 1.19 for trapped particles. The absorbed doses obtained by the RRMD-III and a conventional method using TLD (Mg2SiO4), which was placed around the RRMD-III were compared. It was found that the TLDs showed a lower efficiency, just 58% of absorbed dose registered by the RRMD-III. Copyright (C) 1999 Elsevier Science B.V.
KW - Double-sided silicon detector
KW - LET distribution
KW - Radiation dosimeter
KW - Real time
KW - Space Shuttle
KW - TLD
UR - http://www.scopus.com/inward/record.url?scp=0033394656&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0033394656&partnerID=8YFLogxK
U2 - 10.1016/S0027-5107(99)00130-X
DO - 10.1016/S0027-5107(99)00130-X
M3 - Article
C2 - 10631333
AN - SCOPUS:0033394656
SN - 0027-5107
VL - 430
SP - 191
EP - 202
JO - Mutation Research - Fundamental and Molecular Mechanisms of Mutagenesis
JF - Mutation Research - Fundamental and Molecular Mechanisms of Mutagenesis
IS - 2
ER -