T1 - Effects of Mechanical Stress on Corrosion Resistance of Metallic Biomaterials
AU - Hisamori, Noriyuki
AU - Komotori, Jun
AU - Nakajima, Akiko
N1 - Publisher Copyright:
© Ti-2007 - Proceedings of the 11th World Conference on Titanium. All Rights Reserved.
PY - 2007
Y1 - 2007
N2 - Metallic biomaterials are commonly used as bone fixation devices in the field of orthopedics. They are usually used under a complex stress field in vivo. For example, they are statically stressed by the weight of the patient, and dynamically by exercise of the patient. Furthermore, in some cases, they are plastically strained and adjusted to the shape of the affected part. These stresses may degrade mechanical strength of the implanted metals. Implanted metallic biomaterials tend to corrode over time. Corrosion of implanted metals is harmful because it also causes strength degradation. Moreover, released metallic ions may be toxic to humans. In this study, in order to clarify the influence of the stress and pre-strain (10% and 20%) and their combination effect on corrosion behavior of AISI 316L stainless steel and Ti-6Al-4V alloy in vitro, a new testing system was developed. The system can apply load on a specimen in corrosion fluid. An electrochemical process was used to accelerate corrosion and to measure corrosion damage quantitatively. Anode polarization test were carried out in Ringer's solution. Corrosion resistance was assessed by measuring electric charge densities. After the test, the specimens were observed with a digital microscope and scanning electron microscope. Significant effects of stress and pre-strain on corrosion behavior were observed.
AB - Metallic biomaterials are commonly used as bone fixation devices in the field of orthopedics. They are usually used under a complex stress field in vivo. For example, they are statically stressed by the weight of the patient, and dynamically by exercise of the patient. Furthermore, in some cases, they are plastically strained and adjusted to the shape of the affected part. These stresses may degrade mechanical strength of the implanted metals. Implanted metallic biomaterials tend to corrode over time. Corrosion of implanted metals is harmful because it also causes strength degradation. Moreover, released metallic ions may be toxic to humans. In this study, in order to clarify the influence of the stress and pre-strain (10% and 20%) and their combination effect on corrosion behavior of AISI 316L stainless steel and Ti-6Al-4V alloy in vitro, a new testing system was developed. The system can apply load on a specimen in corrosion fluid. An electrochemical process was used to accelerate corrosion and to measure corrosion damage quantitatively. Anode polarization test were carried out in Ringer's solution. Corrosion resistance was assessed by measuring electric charge densities. After the test, the specimens were observed with a digital microscope and scanning electron microscope. Significant effects of stress and pre-strain on corrosion behavior were observed.
KW - corrosion resistance
KW - pre-strain
KW - stress
KW - surface roughness
KW - SUS316L
KW - Ti-6Al-4V alloy
UR - https://www.scopus.com/pages/publications/85182284139
UR - https://www.scopus.com/pages/publications/85182284139#tab=citedBy
M3 - Conference contribution
AN - SCOPUS:85182284139
T3 - Ti-2007 - Proceedings of the 11th World Conference on Titanium
SP - 1221
EP - 1224
BT - Ti-2007 - Proceedings of the 11th World Conference on Titanium
A2 - Niinomi, M.
A2 - Akiyama, S.
A2 - Ikeda, M.
A2 - Hagiwara, M.
A2 - Maruyama, K.
PB - Japan Institute of Metals (JIM)
T2 - 11th World Conference on Titanium, Ti-2007
Y2 - 3 June 2007 through 7 June 2007
ER -