Abstract
An implantable artificial kidney can markedly improve the quality of life of renal disease patients. Our group has developed an implantable multilayered dialysis system consisting of microfluidic channels and dialysis membranes. Long-term evaluation is necessary for implant devices where biofouling is a critical factor, culminating in the deterioration of dialysis performance. Our previous work revealed that surface conditions, which depend on the manufacturing process, determine the amount of biofouling, and that electrolytic etching is the most suitable technique for forming a channel wall free of biofouling. In this study, we investigated the electrolytic etching conditions in detail. We conducted in vitro experiments for 7 d and evaluated the adhesion of biomaterials by scanning electron microscopy. The experiments revealed that a surface mirrorfinished by electrolytic etching effectively prevents biofouling.
| Original language | English |
|---|---|
| Article number | 06GN10 |
| Journal | Japanese journal of applied physics |
| Volume | 56 |
| Issue number | 6 |
| DOIs | |
| Publication status | Published - 2017 Jun |
ASJC Scopus subject areas
- General Engineering
- General Physics and Astronomy
Fingerprint
Dive into the research topics of 'Evaluation of biofouling in stainless microfluidic channels for implantable multilayered dialysis device'. Together they form a unique fingerprint.Cite this