TY - JOUR
T1 - Evaluation of new immunochromatographic assay kit for adenovirus detection in throat swab
T2 - Comparison with culture and real-time PCR results
AU - Morozumi, Miyuki
AU - Shimizu, Hideaki
AU - Matsushima, Yuki
AU - Mitamura, Keiko
AU - Tajima, Takeshi
AU - Iwata, Satoshi
AU - Ubukata, Kimiko
PY - 2014/5
Y1 - 2014/5
N2 - A new immunochromatographic (IC) assay kit, BD Veritor System Adeno was evaluated to comparing with commercial available kit, BD Adeno Examan, cell culture, and real-time PCR using throat swab samples. Specimens were collected from 146 pediatric patients between July 2011 and January 2012. Mean age of patients was 4 years (8 monthse15 years old). Patients were diagnosed with pharyngitis (n = 67), tonsillitis (n = 45), pharyngoconjunctival fever (n = 26), upper respiratory tract infection (n = 6), conjunctivitis (n = 1), or bronchitis (n = 1). Thirty-one of the patients (21.2%) had more than one disease. Among all samples, 61 (41.8%) were positive for adenovirus with BD Veritor System Adeno; 68 (46.6%) with BD Adeno Examan; 63 (43.2%) with real-time PCR; and 65 (44.5%) with cell culture. Serotype 3 (n = 41; 63.1%) was predominant among the 65 adenovirus isolates, followed by serotype 2 (n = 12; 18.5%), 1 (n = 6; 9.2%), 5 (n = 4; 6.2%), and 4 (n = 2; 3.1%). Relative sensitivity and specificity of BD Veritor System Adeno, BD Adeno Examan, and real-time PCR were 93.8% and 98.7%, 96.9% and 93.8%, and 96.9% and 100%, respectively. Positive predictive and negative predictive values for these methods were 98.4% and 95.1%, 92.6% and 97.4%, and 100% and 97.6%, respectively. The sensitivity and specificity of real-time PCR was greater than that of IC assay kits. However, IC assay kits also showed high sensitivity and specificity appropriate for clinical use.
AB - A new immunochromatographic (IC) assay kit, BD Veritor System Adeno was evaluated to comparing with commercial available kit, BD Adeno Examan, cell culture, and real-time PCR using throat swab samples. Specimens were collected from 146 pediatric patients between July 2011 and January 2012. Mean age of patients was 4 years (8 monthse15 years old). Patients were diagnosed with pharyngitis (n = 67), tonsillitis (n = 45), pharyngoconjunctival fever (n = 26), upper respiratory tract infection (n = 6), conjunctivitis (n = 1), or bronchitis (n = 1). Thirty-one of the patients (21.2%) had more than one disease. Among all samples, 61 (41.8%) were positive for adenovirus with BD Veritor System Adeno; 68 (46.6%) with BD Adeno Examan; 63 (43.2%) with real-time PCR; and 65 (44.5%) with cell culture. Serotype 3 (n = 41; 63.1%) was predominant among the 65 adenovirus isolates, followed by serotype 2 (n = 12; 18.5%), 1 (n = 6; 9.2%), 5 (n = 4; 6.2%), and 4 (n = 2; 3.1%). Relative sensitivity and specificity of BD Veritor System Adeno, BD Adeno Examan, and real-time PCR were 93.8% and 98.7%, 96.9% and 93.8%, and 96.9% and 100%, respectively. Positive predictive and negative predictive values for these methods were 98.4% and 95.1%, 92.6% and 97.4%, and 100% and 97.6%, respectively. The sensitivity and specificity of real-time PCR was greater than that of IC assay kits. However, IC assay kits also showed high sensitivity and specificity appropriate for clinical use.
KW - Adenovirus
KW - Immunochromatographic assay
KW - Rapid detection
KW - Real-time PCR
KW - Serotype
UR - http://www.scopus.com/inward/record.url?scp=84903731562&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84903731562&partnerID=8YFLogxK
U2 - 10.1016/j.jiac.2014.01.005
DO - 10.1016/j.jiac.2014.01.005
M3 - Article
C2 - 24594452
AN - SCOPUS:84903731562
SN - 1341-321X
VL - 20
SP - 303
EP - 306
JO - Journal of Infection and Chemotherapy
JF - Journal of Infection and Chemotherapy
IS - 5
ER -