Abstract
A filament driven multi-cusp negative ion source has been developed for proton cyclotrons in medical applications. In Cs-free operation, continuous H- beam of 10 mA and D- beam of 3.3 mA were obtained stably at an arc-discharge power of 3 kW and 2.4 kW, respectively. In Cs-seeded operation, H- beam current reached 22 mA at a lower arc power of 2.6 kW with less co-extracted electron current. The optimum gas flow rate, which gives the highest H- current, was 15 sccm in the Cs-free operation, while it decreased to 4 sccm in the Cs-seeded operation. The relationship between H- production and the design/operating parameters has been also investigated by a numerical study with KEIO-MARC code, which gives a reasonable explanation to the experimental results of the H- current dependence on the arc power.
| Original language | English |
|---|---|
| Article number | 02B135 |
| Journal | Review of Scientific Instruments |
| Volume | 87 |
| Issue number | 2 |
| DOIs | |
| Publication status | Published - 2016 Feb 1 |
ASJC Scopus subject areas
- Instrumentation
Fingerprint
Dive into the research topics of 'High current DC negative ion source for cyclotron'. Together they form a unique fingerprint.Cite this