TY - JOUR
T1 - In Vivo Metacarpophalangeal Joint Kinematics After Silicone Implant Arthroplasty in Patients With Rheumatoid Arthritis
AU - Ishii, Kazunori
AU - Iwamoto, Takuji
AU - Oki, Satoshi
AU - Nagura, Takeo
AU - Matsumoto, Morio
AU - Nakamura, Masaya
N1 - Publisher Copyright:
© 2023 American Society for Surgery of the Hand
PY - 2023/1
Y1 - 2023/1
N2 - Purpose: The aim of this study was to determine the potential mechanism of implant fracture using 3-dimensional motion analysis of patients with rheumatoid arthritis. Methods: Active flexion motion in 9 hands (34 fingers) of 6 female patients with rheumatoid arthritis who previously underwent hinged silicone metacarpophalangeal joint arthroplasty was examined using 4-dimensional computed tomography. Positions of the proximal phalanges relative to the metacarpals were quantified using a surface registration method. The deformation of the silicone implant was classified in the sagittal plane in the maximum flexion frame. The longitudinal bone axis of the proximal phalanx and the helical axis of the proximal phalanx were evaluated in 3-dimensional coordinates based on the hinge of the silicone implant. Results: Nineteen fingers were classified into group 1, in which the silicone implant moved volarly during flexion without buckling of the distal stem. Twelve fingers were classified into group 2, in which the distal stem of the silicone implant buckled. Three fingers were classified into group 3, in which the base of the distal stem had already fractured. Quantitatively, the longitudinal bone axes of the proximal phalanges were displaced from dorsal to volar in the middle stage of flexion and migrated in the proximal direction in the late phase of flexion. The helical axes of the proximal phalanges were located on the dorsal and proximal sides of the hinge, and these tended to move in the volar and proximal directions as the metacarpophalangeal joint flexed. Conclusions: Volar and proximal translation of the proximal phalange was observed on 4-dimensional computed tomography. Clinical relevance: Proximal displacement of the bone axis late in flexion appears to be a contributing factor inducing implant fractures, because the pistoning motion does not allow the implant to move in the proximal direction.
AB - Purpose: The aim of this study was to determine the potential mechanism of implant fracture using 3-dimensional motion analysis of patients with rheumatoid arthritis. Methods: Active flexion motion in 9 hands (34 fingers) of 6 female patients with rheumatoid arthritis who previously underwent hinged silicone metacarpophalangeal joint arthroplasty was examined using 4-dimensional computed tomography. Positions of the proximal phalanges relative to the metacarpals were quantified using a surface registration method. The deformation of the silicone implant was classified in the sagittal plane in the maximum flexion frame. The longitudinal bone axis of the proximal phalanx and the helical axis of the proximal phalanx were evaluated in 3-dimensional coordinates based on the hinge of the silicone implant. Results: Nineteen fingers were classified into group 1, in which the silicone implant moved volarly during flexion without buckling of the distal stem. Twelve fingers were classified into group 2, in which the distal stem of the silicone implant buckled. Three fingers were classified into group 3, in which the base of the distal stem had already fractured. Quantitatively, the longitudinal bone axes of the proximal phalanges were displaced from dorsal to volar in the middle stage of flexion and migrated in the proximal direction in the late phase of flexion. The helical axes of the proximal phalanges were located on the dorsal and proximal sides of the hinge, and these tended to move in the volar and proximal directions as the metacarpophalangeal joint flexed. Conclusions: Volar and proximal translation of the proximal phalange was observed on 4-dimensional computed tomography. Clinical relevance: Proximal displacement of the bone axis late in flexion appears to be a contributing factor inducing implant fractures, because the pistoning motion does not allow the implant to move in the proximal direction.
KW - Four-dimensional computed tomography
KW - metacarpophalangeal joint
KW - rheumatoid arthritis
KW - silicone arthroplasty
UR - http://www.scopus.com/inward/record.url?scp=85119617312&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85119617312&partnerID=8YFLogxK
U2 - 10.1016/j.jhsa.2021.10.001
DO - 10.1016/j.jhsa.2021.10.001
M3 - Article
C2 - 34823922
AN - SCOPUS:85119617312
SN - 0363-5023
VL - 48
SP - 88.e1-88.e11
JO - Journal of Hand Surgery
JF - Journal of Hand Surgery
IS - 1
ER -