Abstract
Evaluation of the initial stability of implants is essential to reduce the number of implant failures of pedicle screws after orthopedic surgeries. Laser resonance frequency analysis (L-RFA) has been recently proposed as a viable diagnostic scheme in this regard. In a previous study, L-RFA was used to demonstrate the diagnosis of implant stability of monoaxial screws with a fixed head. However, polyaxial screws with movable heads are also frequently used in practice. In this paper, we clarify the characteristics of the laser-induced vibrational spectra of polyaxial screws which are required for making L-RFA diagnoses of implant stability. In addition, a novel analysis scheme of a vibrational spectrum using L-RFA based on machine learning is demonstrated and proposed. The proposed machine learning-based diagnosis method demonstrates a highly accurate prediction of implant stability (peak torque) for polyaxial pedicle screws. This achievement will contribute an important analytical method for implant stability diagnosis using L-RFA for implants with moving parts and shapes used in various clinical situations.
Original language | English |
---|---|
Article number | 7553 |
Journal | Sensors |
Volume | 21 |
Issue number | 22 |
DOIs | |
Publication status | Published - 2021 Nov 1 |
Keywords
- Laser resonance frequency analysis
- Orthopedics
- Pedicle screw
- Stability diagnosis
ASJC Scopus subject areas
- Analytical Chemistry
- Information Systems
- Atomic and Molecular Physics, and Optics
- Biochemistry
- Instrumentation
- Electrical and Electronic Engineering