Abstract
The ion selectivities for poly(vinyl chloride) (PVC)-matrix membrane ion-selective electrodes based on seven lipophilic diketone compounds were examined. All the electrodes based on the tested five β-diketone cdmpounds exhibited clear Mg2. - selectivity under high pH conditions(pH10). Especially, the electrode based on 2-acetyl-l-tetralone shows excellent Mg2. -se1ectivity, and the selectivity coefficients of the Mg2+ -electrode to Ca2+, K+ and Na+(log [formula omitted] log [formula omitted] and log [formula omitted]) in pH 10 solution were -2.2, -1.9 and -2.7, respectively. This electrode gave a response slope of about 60 mV/pMg2+ ranging from 1 Χ 10-4 to 2 Χ 10-2 M Mg2+, which corresponds to twice the usual Nernstian theoretical response slope for the divalent cation. This phenomenon is advantageous for the accurate determination of Mg2+.
Original language | English |
---|---|
Pages (from-to) | 1993-2004 |
Number of pages | 12 |
Journal | Analytical Letters |
Volume | 23 |
Issue number | 11 |
DOIs | |
Publication status | Published - 1990 Nov |
Keywords
- diketone
- ion-selective electrode
- magnesium
- selectivity
ASJC Scopus subject areas
- Analytical Chemistry
- Biochemistry
- Spectroscopy
- Clinical Biochemistry
- Biochemistry, medical
- Electrochemistry