Skip to main navigation Skip to search Skip to main content

Measurement of beam-axis displacement from the isocenter during three-dimensional conformal radiosurgery with a micro-multileaf collimator

  • Etsuo Kunieda
  • , Osamu Kawaguchi
  • , Hidetoshi Saitoh
  • , Tatsuya Fujisaki
  • , Atsuya Takeda
  • , Takatsugu Kawase
  • , Hossain M. Deloar
  • , Naoyuki Shigematsu
  • , Atsushi Kubo

Research output: Contribution to journal › Article › peer-review

Abstract

We describe the displacement of the beam-axis from the planning isocenter in clinical situations during three-dimensional conformal radiosurgery using an Acculeaf bi-directional micro-multileaf collimator. The displacements were recorded for 64 ports using a video imaging system and a stereotactic arc. The mean displacement was 0.41±0.25 mm.

Original languageEnglish
Pages (from-to)45-48
Number of pages4
JournalRadiotherapy and Oncology
Volume70
Issue number1
DOIs
Publication statusPublished - 2004 Jan

Keywords

  • Accuracy
  • Beam's eye monitor
  • Linear accelerator
  • Radiosurgery

ASJC Scopus subject areas

  • Hematology
  • Oncology
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Measurement of beam-axis displacement from the isocenter during three-dimensional conformal radiosurgery with a micro-multileaf collimator'. Together they form a unique fingerprint.

Cite this