Abstract
We describe the displacement of the beam-axis from the planning isocenter in clinical situations during three-dimensional conformal radiosurgery using an Acculeaf bi-directional micro-multileaf collimator. The displacements were recorded for 64 ports using a video imaging system and a stereotactic arc. The mean displacement was 0.41±0.25 mm.
| Original language | English |
|---|---|
| Pages (from-to) | 45-48 |
| Number of pages | 4 |
| Journal | Radiotherapy and Oncology |
| Volume | 70 |
| Issue number | 1 |
| DOIs | |
| Publication status | Published - 2004 Jan |
Keywords
- Accuracy
- Beam's eye monitor
- Linear accelerator
- Radiosurgery
ASJC Scopus subject areas
- Hematology
- Oncology
- Radiology Nuclear Medicine and imaging
Fingerprint
Dive into the research topics of 'Measurement of beam-axis displacement from the isocenter during three-dimensional conformal radiosurgery with a micro-multileaf collimator'. Together they form a unique fingerprint.Cite this
- APA
- Standard
- Harvard
- Vancouver
- Author
- BIBTEX
- RIS