Microscopic Laser Doppler Velocimeter Measuring Blood Velocity in Single Microvessel

Haruyuki Minamitani, Eiji Okada

Research output: Contribution to journalArticlepeer-review

1 Citation (Scopus)

Abstract

A microscopic laser Doppler velocimeter to measure blood velocity in single microvessels was developed. In order to measure the blood velocity in microvessels correctly, the influence of stray scattered light from vessel wall should be removed. Because it provokes low-frequency spurious components. The spectra of the spurious components overlap with that of Doppler signal and induces considerable error in the result of the blood velocity when the velocity is reduced. The optical configuration of the velocimeter was optimized to eliminate the influence of the stray scattered light. The blood velocity in microvessels in rat mesentery was measured with the microscopic laser Doppler velocimeter and the periodical fluctuation of the velocity was analyzed. Pulsatile velocity fluctuation could be observed even in capillary.

Original languageEnglish
Pages (from-to)186-190
Number of pages5
Journalthe keio journal of medicine
Volume42
Issue number4
DOIs
Publication statusPublished - 1993