TY - JOUR
T1 - Platelet deposition onto polymeric surfaces during shunting
AU - Fujimoto, Keiji
AU - Minato, Masao
AU - Tadokoro, Hideki
AU - Ikada, Yoshito
PY - 1993/3
Y1 - 1993/3
N2 - In an attempt to develop blood‐contacting tubes that can be applied for short‐term uses with a reduced heparin concentration or, ideally, without heparinization, we evaluated the blood compatibility of polymeric materials with a rabbit ex vivo shunt model. The shunt tubes employed were made of silicone, plasticized poly(vinyl chloride) (PVC), and segmented poly(ether urethane) (PU). In addition, two kinds of surface‐modified tube were used: poly(vinyl alcohol) (PVA)‐coated PVC and poly(dimethylacrylamide) (PDMAA)‐grafted PU. The ex vivo shunt results correlated well with protein adsorption and platelet adhesion in vitro. The following order for the extent of platelet deposition was given, irrespective of the blood‐contacting duration: PDMAA‐grafted PU < PVA‐coated PVC < PU < silicone, PVC. It is likely that many platelet aggregates detached from the PVA‐coated PVC surface. For PDMAA‐grafted PU, no trace of detachment of aggregates could be detected on any of the SEM photographs. The number and morphology of blood cells adhered onto the tube surfaces during ex vivo shunting were dependent on the kind of polymer surfaces, the blood exposure time, and the flow rate of blood. © 1993 John Wiley & Sons, Inc.
AB - In an attempt to develop blood‐contacting tubes that can be applied for short‐term uses with a reduced heparin concentration or, ideally, without heparinization, we evaluated the blood compatibility of polymeric materials with a rabbit ex vivo shunt model. The shunt tubes employed were made of silicone, plasticized poly(vinyl chloride) (PVC), and segmented poly(ether urethane) (PU). In addition, two kinds of surface‐modified tube were used: poly(vinyl alcohol) (PVA)‐coated PVC and poly(dimethylacrylamide) (PDMAA)‐grafted PU. The ex vivo shunt results correlated well with protein adsorption and platelet adhesion in vitro. The following order for the extent of platelet deposition was given, irrespective of the blood‐contacting duration: PDMAA‐grafted PU < PVA‐coated PVC < PU < silicone, PVC. It is likely that many platelet aggregates detached from the PVA‐coated PVC surface. For PDMAA‐grafted PU, no trace of detachment of aggregates could be detected on any of the SEM photographs. The number and morphology of blood cells adhered onto the tube surfaces during ex vivo shunting were dependent on the kind of polymer surfaces, the blood exposure time, and the flow rate of blood. © 1993 John Wiley & Sons, Inc.
UR - http://www.scopus.com/inward/record.url?scp=0027572385&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0027572385&partnerID=8YFLogxK
U2 - 10.1002/jbm.820270307
DO - 10.1002/jbm.820270307
M3 - Article
C2 - 8360203
AN - SCOPUS:0027572385
SN - 0021-9304
VL - 27
SP - 335
EP - 343
JO - Journal of Biomedical Materials Research
JF - Journal of Biomedical Materials Research
IS - 3
ER -