Abstract
Purpose : To evaluate the reliability of Pentacam® in the measurement of central corneal thickness. Cases and Method : This prospective study was made on 116 eyes of 116 patients with primary open-angle glaucoma. The age ranged from 35 to 82 years, average 61 years. Thickness of central cornea was measured by Pentacam®, ultrasonography and noncontact specular microscopy. Results : Central corneal thickness averaged 534.3±35.6μm by Pentacam®, 526.5±33.9μm by ultrasonography, and 512.7± 38.7 μm by specular microscopy. These three values were significantly correlated. No significant difference was present between values by Pentacam® and ultrasonography (p=0.30). Values by Pentacam® and specular microscopy were significantly different (p<0.00l). Conclusion : Pentacam® shows same values for central corneal thickness as ultrasonography in eyes with open-angle glaucoma.
Original language | English |
---|---|
Pages (from-to) | 1777-1781 |
Number of pages | 5 |
Journal | Japanese Journal of Clinical Ophthalmology |
Volume | 63 |
Issue number | 12 |
Publication status | Published - 2009 Nov |
ASJC Scopus subject areas
- Ophthalmology