Transcutaneous blood glucose monitoring system based on an ISFET glucose sensor and studies on diabetic patients.

N. Ito, A. Saito, S. Kayashima, J. Kimura, T. Kuriyama, N. Nagata, T. Arai, M. Kikuchi

Research output: Contribution to journalArticlepeer-review

7 Citations (Scopus)

Abstract

A transcutaneous blood glucose monitoring system consists of an ion-sensitive field-effect transistor (ISFET) glucose sensor unit and a suction effusion fluid (SEF) collecting unit. The SEF is directly collected by a weak suction (400 mmHg absolute pressure) through the skin from which the corneum layer of the epidermis has been previously removed. An ISFET glucose sensor unit is able to measure glucose concentrations in a microliter order sampling volume. The system was applied to three diabetic patients during a 75 g oral glucose tolerance test for monitoring blood glucose levels. During the experiments, glucose changes in the SEF followed actual blood glucose levels with 10 min delays. Results suggest the feasibility of utilizing quasi-continuous, transcutaneous blood glucose monitoring for individual patients with various diabetic histories or diabetic complications.

Original languageEnglish
Pages (from-to)269-280
Number of pages12
JournalFrontiers of medical and biological engineering : the international journal of the Japan Society of Medical Electronics and Biological Engineering
Volume6
Issue number4
Publication statusPublished - 1995
Externally publishedYes

ASJC Scopus subject areas

  • Biophysics

Fingerprint

Dive into the research topics of 'Transcutaneous blood glucose monitoring system based on an ISFET glucose sensor and studies on diabetic patients.'. Together they form a unique fingerprint.