TY - JOUR
T1 - Ultrafast 1D MR thermometry using phase or frequency mapping
AU - Mei, Chang Sheng
AU - Mulkern, Robert V.
AU - Oshio, Koichi
AU - Chen, Nan Kuei
AU - Madore, Bruno
AU - Panych, Lawrence P.
AU - Hynynen, Kullervo
AU - McDannold, Nathan J.
N1 - Funding Information:
Acknowledgments This work was supported by NIH grants R01HL077606, U41RR019703, and P01CA067165.
PY - 2012/2
Y1 - 2012/2
N2 - Object To develop an ultrafastMRI-based temperaturemonitoring method for application during rapid ultrasound exposures in moving organs. Materials and methods A slice selective 90? ? 180? pair of RF pulses was used to solicit an echo from a column, which was then sampled with a train of gradient echoes. In a gel phantom, phase changes of each echo were compared to standard gradient-echo thermometry, and temperature monitoring was tested during focused ultrasound sonications. Signal-to-noise ratio (SNR) performance was evaluated in vivo in a rabbit brain, and feasibility was tested in a human heart. Results The correlation between each echo in the acquisition and MRI-based temperature measurements was good (R = 0.98 ± 0.03). A temperature sampling rate of 19Hzwas achieved at 3T in the gel phantom. It was possible to acquire the water frequency in the beating heart muscle with 5-Hz sampling rate during a breath hold. Conclusion Ultrafast thermometry via phase or frequency monitoring along single columns was demonstrated. With a temporal resolution around 50 ms, it may be possible to monitor focal heating produced by short ultrasound pulses.
AB - Object To develop an ultrafastMRI-based temperaturemonitoring method for application during rapid ultrasound exposures in moving organs. Materials and methods A slice selective 90? ? 180? pair of RF pulses was used to solicit an echo from a column, which was then sampled with a train of gradient echoes. In a gel phantom, phase changes of each echo were compared to standard gradient-echo thermometry, and temperature monitoring was tested during focused ultrasound sonications. Signal-to-noise ratio (SNR) performance was evaluated in vivo in a rabbit brain, and feasibility was tested in a human heart. Results The correlation between each echo in the acquisition and MRI-based temperature measurements was good (R = 0.98 ± 0.03). A temperature sampling rate of 19Hzwas achieved at 3T in the gel phantom. It was possible to acquire the water frequency in the beating heart muscle with 5-Hz sampling rate during a breath hold. Conclusion Ultrafast thermometry via phase or frequency monitoring along single columns was demonstrated. With a temporal resolution around 50 ms, it may be possible to monitor focal heating produced by short ultrasound pulses.
KW - Echo-planar magnetic resonance imaging
KW - MR spectroscopy
KW - Proton resonance frequency shift
KW - Thermometry
UR - http://www.scopus.com/inward/record.url?scp=84862643083&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84862643083&partnerID=8YFLogxK
U2 - 10.1007/s10334-011-0272-9
DO - 10.1007/s10334-011-0272-9
M3 - Article
C2 - 21800192
AN - SCOPUS:84862643083
SN - 0968-5243
VL - 25
SP - 5
EP - 14
JO - Magnetic Resonance Materials in Physics, Biology and Medicine
JF - Magnetic Resonance Materials in Physics, Biology and Medicine
IS - 1
ER -