Abnormally high Patient State Index associated with epicardial pacing: a case report

  • Hiroki Yonezawa
  • , Takuma Maeda
  • , Yoshiaki Takise
  • , Masahiro Morinaga
  • , Yoshihiko Ohnishi

研究成果: Article › 査読

2   !!Link opens in a new tab 被引用数 (Scopus)

抄録

Background: Anesthesiologists monitor electroencephalography (EEG) intraoperatively to maintain adequate depth of anesthesia. However, the EEG signal is affected by noise and interference. The SedLine® is a brain function monitor with which the Patient State Index (PSI) is calculated. In this study, we report abnormally high PSI values associated with epicardial pacing during open heart surgery. Case presentation: A 50-year-old man was scheduled for total arch replacement. Atrial demand pacing was started before weaning from cardiopulmonary bypass. The PSI increased from 30 to 80 soon after the start of pacing, and the EEG waveform showed spikes synchronized with the pacing. As the pacing output was lowered, the spikes on the EEG attenuated and disappeared, and the PSI decreased to < 40. When the pacing output was increased again, the spikes recurred, and the PSI increased again. Conclusions: Pacemaker spikes may cause contamination of the EEG, resulting in abnormally high PSI values.