Accuracy of MR temperature measurement based on chemical shift change for radiofrequency ablation using hook-shaped electrodes

Shigeo Okuda, Kagayaki Kuroda, Osamu Kainuma, Koichi Oshio, Hirokazu Fujiwara, Sachio Kuribayashi

研究成果: Article査読

7 被引用数 (Scopus)

抄録

The purpose of the current study is to evaluate the accuracy of MR thermometry for radiofrequency ablation (RFA) with hook-shaped electrodes. The objects were eight extracted bovine livers. The chemical shift change was calculated from MR images acquired with a spoiled gradient echo sequence and compared with the temperature directly measured with a thermocouple. Linear regression was established between them with a coefficient of -0.0110±0.0007 ppm/°C and errors were calculated as -0.50±7.50°C. MR thermometry is capable of monitoring temperature for RFA.

本文言語English
ページ(範囲)95-100
ページ数6
ジャーナルMagnetic Resonance in Medical Sciences
3
2
DOI
出版ステータスPublished - 2004