TY - JOUR
T1 - Current status and issues with the dosimetric assay of iodine-125 seed sources at medical facilities in Japan
T2 - a questionnaire-based survey
AU - Kojima, Toru
AU - Kawamura, Shinji
AU - Otani, Yuki
AU - Hanada, Takashi
AU - Wakitani, Yuichiro
AU - Naniwa, Kenta
AU - Yorozu, Atsunori
AU - Ikushima, Hitoshi
AU - Dokiya, Takushi
N1 - Publisher Copyright:
© 2023 The Author(s). Published by Oxford University Press on behalf of The Japanese Radiation Research Society and Japanese Society for Radiation Oncology.
PY - 2023/11/1
Y1 - 2023/11/1
N2 - In conducting dosimetric assays of seed sources containing iodine-125 (125I), several major guidelines require the medical physicist to verify the source strength before patient treatment. Japanese guidelines do not mandate dosimetric assays at medical facilities, but since 2017, three incidents have occurred in Japan wherein seeds with incorrect strengths were delivered to medical facilities. Therefore, this study aimed to survey the current situation and any barriers to conducting the dosimetric assay of iodine-125 seeds at medical facilities in Japan. We conducted a questionnaire-based survey from December 2020 to April 2021, to examine whether seed assay and verification of the number of seeds delivered were being performed. We found that only 9 facilities (16%) performed seed assay and 28 (52%) verified the number of seeds. None of the facilities used an assay method that ensured traceability. The reasons for not performing an assay were divided into two categories: lack of resources and legal issues. Lack of resources included lack of instruments, lack of knowledge of assay methods, shorthand, or all of the above, whereas legal issues included the inability to resterilize iodine-125 seeds distributed in Japan and/or purchase seeds dedicated to the assay. Dosimetric assays, including simple methods, are effective in detecting calibration date errors and non-radioactive seeds. The study findings suggest that familiarization of medical personnel with these assay methods and investigation of the associated costs of labor and equipment should be recommended, as these measures will lead to medical reimbursement for quality assurance.
AB - In conducting dosimetric assays of seed sources containing iodine-125 (125I), several major guidelines require the medical physicist to verify the source strength before patient treatment. Japanese guidelines do not mandate dosimetric assays at medical facilities, but since 2017, three incidents have occurred in Japan wherein seeds with incorrect strengths were delivered to medical facilities. Therefore, this study aimed to survey the current situation and any barriers to conducting the dosimetric assay of iodine-125 seeds at medical facilities in Japan. We conducted a questionnaire-based survey from December 2020 to April 2021, to examine whether seed assay and verification of the number of seeds delivered were being performed. We found that only 9 facilities (16%) performed seed assay and 28 (52%) verified the number of seeds. None of the facilities used an assay method that ensured traceability. The reasons for not performing an assay were divided into two categories: lack of resources and legal issues. Lack of resources included lack of instruments, lack of knowledge of assay methods, shorthand, or all of the above, whereas legal issues included the inability to resterilize iodine-125 seeds distributed in Japan and/or purchase seeds dedicated to the assay. Dosimetric assays, including simple methods, are effective in detecting calibration date errors and non-radioactive seeds. The study findings suggest that familiarization of medical personnel with these assay methods and investigation of the associated costs of labor and equipment should be recommended, as these measures will lead to medical reimbursement for quality assurance.
KW - iodine-125
KW - low-dose-rate brachytherapy
KW - low-energy source
KW - prostate cancer
KW - seed assay
UR - http://www.scopus.com/inward/record.url?scp=85178495788&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85178495788&partnerID=8YFLogxK
U2 - 10.1093/jrr/rrad063
DO - 10.1093/jrr/rrad063
M3 - Article
C2 - 37697653
AN - SCOPUS:85178495788
SN - 0449-3060
VL - 64
SP - 962
EP - 966
JO - Journal of radiation research
JF - Journal of radiation research
IS - 6
ER -