抄録
This paper proposes a new method which removes artifacts due to T2 relaxation in Rapid Acquisition with Relaxation Enhancement (RARE) images. The method removes the artifacts by dividing the average of two Hermitian symmetric data points in k-space by a hyperbolic cosine function prior to inverse Fourier transformation. In order to evaluate the effectiveness of the method, computer simulations were done in which T2 distribution was spatially variant. The results showed that the proposed method worked well for spatially variant T2 decay and was less sensitive to noise than the conventional method using the geometric mean of two symmetric points and that using an exponential filter. In addition, the method yielded good results for a wide range of T2 values.
| 本文言語 | English |
|---|---|
| ページ | 1457-1460 |
| ページ数 | 4 |
| 出版ステータス | Published - 1995 |
| 外部発表 | はい |
| イベント | Proceedings of the 1994 Nuclear Science Symposium and Medical Imaging Conference. Part 1 (of 4) - Norfolk, VA, USA 継続期間: 1994 10月 30 → 1994 11月 5 |
Other
| Other | Proceedings of the 1994 Nuclear Science Symposium and Medical Imaging Conference. Part 1 (of 4) |
|---|---|
| City | Norfolk, VA, USA |
| Period | 94/10/30 → 94/11/5 |