TY - JOUR
T1 - Is Medical Linac Suitable for High-precision Stereotactic Irradiation?
T2 - Investigations in geometrical accuracies of gantry and couch
AU - Kunieda, Etsuo
AU - Kitamura, Masayuki
AU - Kawaguchi, Osamu
AU - Ohira, Takayuki
AU - Shigematsu, Naoyuki
AU - Tonai, Takenori
AU - Ando, Yutaka
AU - Kubo, Atsushi
AU - Kawase, Takeshi
PY - 1998
Y1 - 1998
N2 - Linac-based radiosurgery has many advantages over the gamma knife, including low initial cost and no need of source replacement. On the other hand, most of the medical linacs currently in use were not originally designed to be applied for radiosurgery, and, therefore, careful quality assurance programs are required. In the gantry-head of a linac, a small CCD video camera is mounted in a position optically identical to that of the x-ray source. The video signal from the camera was digitalized to be evaluated for gemetrical errors. A metal ball fixed to the stereotactic base frame via XYZ-sliding rods was used as a simulated target. Displacements of the target from the isocenter were measured during rotation of the gantry. Displacements in the gantry-rotation plane were satisfactorily small, while those perpendicular to it were maximal at gantry position angles of 0° and 180°. This error night be caused by gravitational vending of the heavy gantry head. Although other major errors of the linac were within one millimeter, the center of coach rotation around the isocenter did not coincide with the center of gantry rotation, probably owing to gravitational vending. Special care should be taken when very small collimators are employed.
AB - Linac-based radiosurgery has many advantages over the gamma knife, including low initial cost and no need of source replacement. On the other hand, most of the medical linacs currently in use were not originally designed to be applied for radiosurgery, and, therefore, careful quality assurance programs are required. In the gantry-head of a linac, a small CCD video camera is mounted in a position optically identical to that of the x-ray source. The video signal from the camera was digitalized to be evaluated for gemetrical errors. A metal ball fixed to the stereotactic base frame via XYZ-sliding rods was used as a simulated target. Displacements of the target from the isocenter were measured during rotation of the gantry. Displacements in the gantry-rotation plane were satisfactorily small, while those perpendicular to it were maximal at gantry position angles of 0° and 180°. This error night be caused by gravitational vending of the heavy gantry head. Although other major errors of the linac were within one millimeter, the center of coach rotation around the isocenter did not coincide with the center of gantry rotation, probably owing to gravitational vending. Special care should be taken when very small collimators are employed.
KW - Accuracy
KW - Linac
KW - Quality
KW - Radiosurgery
UR - http://www.scopus.com/inward/record.url?scp=11744301592&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=11744301592&partnerID=8YFLogxK
M3 - Article
C2 - 9558849
AN - SCOPUS:11744301592
SN - 0048-0428
VL - 58
SP - 86
EP - 90
JO - Nippon Acta Radiologica
JF - Nippon Acta Radiologica
IS - 3
ER -