Measurement of beam-axis displacement from the isocenter during three-dimensional conformal radiosurgery with a micro-multileaf collimator

  • Etsuo Kunieda
  • , Osamu Kawaguchi
  • , Hidetoshi Saitoh
  • , Tatsuya Fujisaki
  • , Atsuya Takeda
  • , Takatsugu Kawase
  • , Hossain M. Deloar
  • , Naoyuki Shigematsu
  • , Atsushi Kubo

研究成果: Article › 査読

抄録

We describe the displacement of the beam-axis from the planning isocenter in clinical situations during three-dimensional conformal radiosurgery using an Acculeaf bi-directional micro-multileaf collimator. The displacements were recorded for 64 ports using a video imaging system and a stereotactic arc. The mean displacement was 0.41±0.25 mm.

本文言語English
ページ(範囲)45-48
ページ数4
ジャーナルRadiotherapy and Oncology
巻70
号1
DOI
出版ステータスPublished - 2004 1月