抄録
A temperature sensor integrated on a micro-device for biological experiments requires affordable, rapid and easy thermal calibration. However, such calibration cannot usually be done directly under the microscope, a fact that impedes biological experiments. We present in this paper an inexpensive and rapid method to achieve thermal calibration directly under the microscope. It is based on the use of a thermotropic liquid crystal: the 4-n-octyl-4- cyanobiphenyl (8 CB) exhibiting an isothermal phase change at 313 K that can be monitored optically. We demonstrate the advantages of this method by calibrating the temperature sensor integrated onto a micro-device.
| 本文言語 | English |
|---|---|
| ページ(範囲) | 1600-1602 |
| ページ数 | 3 |
| ジャーナル | Lab on a Chip |
| 巻 | 7 |
| 号 | 11 |
| DOI | |
| 出版ステータス | Published - 2007 |
| 外部発表 | はい |