メインナビゲーションにスキップ 検索にスキップ メインコンテンツにスキップ

On-chip thermal calibration with 8 CB liquid crystal of micro-thermal device

  • Frédéric Gillot
  • , Fabrice Olivier Morin
  • , Hideyuki F. Arata
  • , Régis Guégan
  • , Hajime Tanaka
  • , Hiroyuki Fujita

研究成果: Article › 査読

11   !!Link opens in a new tab 被引用数 (Scopus)

抄録

A temperature sensor integrated on a micro-device for biological experiments requires affordable, rapid and easy thermal calibration. However, such calibration cannot usually be done directly under the microscope, a fact that impedes biological experiments. We present in this paper an inexpensive and rapid method to achieve thermal calibration directly under the microscope. It is based on the use of a thermotropic liquid crystal: the 4-n-octyl-4- cyanobiphenyl (8 CB) exhibiting an isothermal phase change at 313 K that can be monitored optically. We demonstrate the advantages of this method by calibrating the temperature sensor integrated onto a micro-device.

本文言語English
ページ(範囲)1600-1602
ページ数3
ジャーナルLab on a Chip
巻7
号11
DOI
出版ステータスPublished - 2007