@article{0e52cf5ce6154fcf9e8ff45fd3e4bc9e,
title = "Rotational stability and clinical outcomes of a new one piece toric intraocular lens with anchor-wing haptics",
abstract = "Background: To evaluate the safety and efficacy of a new toric intraocular lens (IOL) with anchor-wing haptics. Methods: The new toric IOL with anchor-wing haptics (NS60YT, NIDEK Co., Ltd.) was implanted in eligible patients with age-related cataracts with preoperative corneal astigmatism of 1.0 D or greater at a university hospital and two private hospitals in Japan. The following IOL cylinder powers were evaluated: 1.50 D (NS60YT3), 2.25 D (NS60YT4), 3.00 D (NS60YT5) and 4.50 D (NS60YT7). All patients were assessed out to 12 months postoperatively. The primary endpoint was visual acuity (VA) with spherical addition at 6 months postoperatively, and the primary analysis calculated the proportion of eyes with VA with spherical addition of 0.1 logMAR or better. The magnitude of rotation was compared to the intended axis of IOL implantation at each postoperative examination. Adverse events were evaluated for the safety analysis. Results: This study enrolled 64 eyes of 53 patients. At 6 months postoperatively, for all IOL powers, VA with spherical addition of 0.1 logMAR or better was achieved in 90% [95% confidence interval (CI): 80–96] of eyes. The mean IOL rotation was 5.3 ± 4.3° at 12 months postoperatively. The mean magnitude of rotation ranged from 1.9° to 2.5° between each postoperative examination from 1 day to 12 months. There were no vision-threatening intraoperative or postoperative complications for the duration of the study. Conclusions: The NS60YT IOL remained stable after implantation and was efficacious for treating 1.00 D or greater astigmatism in patients with senile cataracts. Trial registration: This study was registered at ClinicalTrials.gov (NCT03242486) on August 8, 2017 - Retrospectively registered.",
keywords = "Anchor-wing haptics, Cataract, Cataract surgery, Corneal astigmatism, Intraocular lens, Rotational stability, Toric intraocular lens",
author = "Iichiro Sugita and Tomoichiro Ogawa and Kazuo Ichikawa and Takahide Okita and Kazuno Negishi and Tadashi Nakano and Hiroshi Tsuneoka",
note = "Funding Information: Dr. Sugita, Ogawa, Ichikawa, Okita, Nakano, and Tsuneoka report grants and Dr. Negishi report personal fees from NIDEK CO., LTD., during the conduct of the study;. Dr. Sugita reports grants and personal fees from Alcon Japan Ltd., personal fees and non-financial support from Santen Pharmaceutical Co., Ltd., non-financial support from HOYA Corporation, personal fees from Kowa Co., Ltd., grants and personal fees from Senjyu Pharmaceutical Co., Ltd., personal fees from Carl Zeiss Meditec AG, personal fees from Bayer AG, personal fees from Pfizer Inc., personal fees from Nikon Corporation, outside the submitted work;. Dr. Ichikawa reports grants and personal fees from Alcon, grants and personal fees from ZEISS, grants and personal fees from STAAR, personal fees from HOYA, personal fees from SANTEN, personal fees from KY CenterVue, personal fees from Igaku shoin, personal fees from Kowa, personal fees from JFC, personal fees from Novartis, outside the submitted work;. Dr. Negishi reports grants from Fuji Xerox Co, grants from Hitachi Automotive Systems, Ltd., grants from Universal View Co., Ltd., grants and personal fees from Alcon Japan Ltd., grants and personal fees from Santen Pharmaceutical, grants and personal fees from HOYA Corporation, grants and personal fees from AMO Japan K.K., grants from Kowa Company, grants from Kowa Pharmaceutical Company Ltd., grants from Tomey Corporation, personal fees from Senju Pharmaceutical, personal fees from Otsuka Pharmaceutical, personal fees from NIDEK Co., Ltd., personal fees from Carl Zeiss Meditec AG, outside the submitted work; Dr. Negishi has a patent (PCT/JP2015/65997, PCT/JP2016/085903) for potential products for myopia suppression pending to Tsubota Laboratory Inc. Publisher Copyright: {\textcopyright} 2022, The Author(s).",
year = "2022",
month = dec,
doi = "10.1186/s12886-021-02240-7",
language = "English",
volume = "22",
journal = "BMC ophthalmology",
issn = "1471-2415",
publisher = "BioMed Central",
number = "1",
}