Suprachoroidal hemorrhage caused by breakage of a 25-gauge cannula

Chieh Jen Chen, Shingo Satofuka, Makoto Inoue, Susumu Ishida, Kei Shinoda, Kazuo Tsubota

研究成果: Article査読

15 被引用数 (Scopus)

抄録

A72-year-old woman had vitreous surgery for epiretinal membrane using the 25-gauge vitrectomy system. During the removal of the cannula at the end of the surgery, half of its tip was noted to be missing. The following day, a severe choroidal detachment associated with a hypotony was found. A second surgery was performed, including drainage of suprachoroidal hemorrhage and choroidal fluid and the removal of the tip of the 25-gauge cannula stuck inside the scterotomy. The retained cannula tip might have acted as a channel allowing vitreous fluid into the suprachoroidal space, resulting in choroidal detachment, hemorrhage, and hypotony.

本文言語English
ページ(範囲)323-324
ページ数2
ジャーナルOphthalmic Surgery Lasers and Imaging
39
4
DOI
出版ステータスPublished - 2008 7月