抄録
An echo-planar spectroscopic imaging method of temperature mapping is proposed. This method is sufficiently faster than the so-called 3D magnetic resonance spectroscopic imaging (3D-MRSI) method and does not require image subtractions, unlike the conventional phase mapping method when an internal reference signal is detectable. The water proton chemical shift measured by using the tissue lipid as an internal reference clearly visualized the temperature change in a porcine liver sample in vitro. It was also demonstrated that the internally referenced echo-planar spectroscopic imaging method could markedly reduce a temperature error caused by a simple, translational motion between scans compared with the phase-mapping method. (C) 2000 Wiley-Liss, Inc.
本文言語 | English |
---|---|
ページ(範囲) | 220-225 |
ページ数 | 6 |
ジャーナル | Magnetic Resonance in Medicine |
巻 | 43 |
号 | 2 |
DOI | |
出版ステータス | Published - 2000 |
外部発表 | はい |
ASJC Scopus subject areas
- 放射線学、核医学およびイメージング